ECONOMICAL
VERTICAL LAMINAR AIR FLOW CABINETS
F L S R

|
Clean air
Working contamination
Laboratory air
![]()
See how it works.
pass your mouse over the image to do it. |

Total product protection
at a very low cost
Ideal for Tissue culture, Media
preparation, Sterile handlings, PCR sample preparation ...
Working principle
An air stream, sterilized by an H14 filter is directes as laminar air
flow towards the working surface ensuring total particle elimination in the
work space, the air is exhausted trough the frontal access, avoiding any
contamination.
Construction:
- Painted steel, covered with corrosion resistant paints
- High efficiency filter type H14 (EN 1822) with an 99.995% MPPS
efficiency
- Washable prefilter with 90% efficiency for 5 µm particles
- Low noise, high capacity motorblower to handle all the air movement
requirement in excess.
- Vertical laminar air flow maintaining a Class 100 (ISO 14644)
environment.
- Control panel to allow the adjustment of, air speed, White light , UV
light, etc...
- Permanent stand by speed to maintain the sterility of the working area
overnight.
- Stainless steel woking surface easyly detachable for cleaning.
- Detachable policarbonate front closure.
Optionals.
- Analogic manometer to check the clogging of the filters
- Vacuum , gaz connections
- 220V conections inside
Dimensions:
| FLSR 60 |
Wide |
High |
Depth |
|
| Overall |
620 |
1000 |
620 |
mm |
| Work area |
615 |
560 |
615 |
mm |
| FLSR 80 |
|
|
|
|
| Overall |
810 |
1000 |
620 |
mm |
| Work area |
795 |
560 |
615 |
mm |
| FLSR 95 |
|
|
|
|
| Overall |
950 |
1000 |
620 |
mm |
| Work area |
945 |
600 |
615 |
mm |
| FLSR 120 |
|
|
|
|
| Overall |
1250 |
1000 |
620 |
mm |
| Work area |
1245 |
600 |
615 |
mm |
| FLSR 150 |
|
|
|
|
| Overall |
1550 |
1000 |
620 |
mm |
| Work area |
1545 |
600 |
615 |
mm |
|